The CR-60-HC Stent Crimping Machine is developed for the crimping of balloon expandable bio absorbable stents.
Specifications
- Stent and balloon positioning outside of crimper head.
- Crimp head can be heated (max.80 °C) and cooled (min.10 °C) in the process cycle.
- Inflation and deflation system to create high stent retention values.
- PLC controlled system.
- The CR-60-HC is built according to the MPT Europe BV high quality standards, resulting in a stable and robust design with minimal cost of ownership.
- A hub-connector is available for inflation and deflation of the product during the crimping cycle to achieve better retention.
- 6.5“ full color touch screen user interface for access to the machine parameters.
Technical Specs
Stent crimping length | 60 mm crimp head. Maximum 45 mm length stents |
Stent type | Biopolymer stents |
Maximum diameter loading | 15 mm (optionally larger sizes possible) |
Minimum closing diameter | 0.2 mm (software limit) |
Temperature range | 10 °C – 80 °C |
Crimp head control | Stepper controlled position & force measurement |
Diameter accuracy | +/- 0.02 mm @ 100 N |
Maximum crimp force | 300 N radial |
Crimping speeds | 0.1 mm/s – 10 mm/s |
Inflation / Deflation | 6 bar max. inflation / -0.8 bar deflation |
Stent positioning | Outside the crimp head (manually / optically) |
GUI | 6.5” colour touch panel |
Control software | Recipe based Password levels (operator, engineer, calibrator) Remote support option Data logging with ethernet possible |
Ethernet | 1x |
USB | NA |
Required air pressure | 7 bar minimum |
Power supply | 240 VAC / 1.2 kW |